Develops, manufactures, and Distributes various types of in-vitro diagnostics and plastic vacuum blood collection tubes. Diagnostics Business of SEKISUI MEDICAL.

To Previous Page
We created this page for medical staff

Haloperidol Kit/Bromperidol Kit
CEDIA-HAL/BRP

Measurement Principle

  • CEDIA (Cloned Enzyme Donor Immunoassay) method(Homogeneous enzyme-linked immuno-solvent assay)
  • Applicable to various types of automated analyzers

Cross-reactivity

  Haloperidol Concentration
(ng/mL)
Crossover rate
(%)
Metabolites Reduced haloperidol 1,000 1.7
4-Fluorophenylpropionic acid 100,000 <0.01
4-Fluorophenylacetic acid 100,000 <0.01
4-Fluorophenylaceturic acid 100,000 <0.01
4-Hydroxy-4-(p-chlorophenyl) piperidine 100,000 <0.01
Similar drugs Bromperidol 40 114
Reduced bromperidol 1,000 0.9
Haloperidol decanoate 1,000 1.2
Benperidol 500 6.2
Moperone 60 50.8
Timiperone 500 4.9
Other drugs Levomepromazine 100,000 <0.01
Chlorpromazine 100,000 <0.01
Zotepine 100,000 <0.01
Risperidone 100,000 <0.01
Biperiden 100,000 <0.01
Trihexyphenidyl hydrochloride 100,000 <0.01
Nitrazepam 20,000 <0.01
  Bromperidol Concentration
(ng/mL)
Crossover rate
(%)
Metabolites Reduced bromperidol 1,000 0.9
4-Fluorophenylpropionic acid 100,000 <0.01
4-Fluorophenylacetic acid 100,000 <0.01
4-Fluorophenylaceturic acid 100,000 <0.01
4-Hydroxy-4-(p-chlorophenyl) piperidine 100,000 <0.01
Similar drugs Haloperidol 40 97.5
Reduced haloperidol 1,000 1.5
Haloperidol decanoate 1,000 1.1
Benperidol 500 5.6
Moperone 60 46.4
Timiperone 500 4.5
Other drugs Levomepromazine 100,000 <0.01
Chlorpromazine 100,000 <0.01
Zotepine 100,000 <0.01
Risperidone 100,000 <0.01
Biperiden 100,000 <0.01
Trihexyphenidyl hydrochloride 100,000 <0.01
Nitrazepam 20,000 <0.01

The structure of haloperidol closely resembles that of bromperidol (the atom at the terminus of the side chain is Cl for haloperidol and Br for bromperidol). Accordingly, both substances have similar reactivities to CEDIA-HAL/BRP, so a sample that contains both haloperidol and bromperidol doesn't provides accurate measured data for each.

To Page Top

Example of Assay Diagram

We have applications suitable for various types of automated analyzers available. For assistance with applications on automated analyzers, please contact us.
Before measurement of haloperidol, perform calibration using a CEDIA haloperidol Calibrator. Before measurement of bromperidol, perform calibration using a CEDIA bromperidol Calibrator.

Measurement range

  • Haloperidol: 1.25-40ng/mL
  • Bromperidol: 1.25-40ng/mL

Examples of Optimum Concentrations in Blood

Therapeutic window

  • Haloperidol: 5.6-16.9ng/mL3)* 5-15ng/mL4) (GC)
  • Bromperidol: 4.0-14.3ng/mL5) (HPLC)

Sufficient concentration by therapeutic plateau

  • Haloperidol: 13ng/mL6) (HPLC)
  • Bromperidol: 12ng/mL7) (HPLC)

*By meta-analysis (HPLC, GC, RIA)

Shelf Life of Reconstituted Reagents

60 days at 2-8°C

References

  • Shinya Sakamoto, et al: Nihon Rinsho, Extra number 57, 380 (1999)
  • Norio Yasui-Furukori: Submitted
  • Ulrich S, et al: Clin Pharmacokinet 34, 227 (1998)
  • Irl Extein, et al: Psychopharmacology bulletin 18, 156 (1982)
  • Tetsuro Takagi, et al: Japanese Journal of Clinical and Experimental Medicine, 66, 2670 (1989)
  • Toshiyuki Someya, et al: Neuropsychopharmacology 18, 511 (1996)
  • Kazuyuki Fujita, et al: Psychiatric therapeutics 16, 581 (2001)
Correlation with HPLC Method
To Previous Page

We prepared this page for medical staff (doctors, dentists, pharmacists, clinical laboratory technicians, nurses, etc.) in Japan to ensure the correct use of our products. You should be aware that the information provided is not intended for overseas medical staff and the general public.