Procalcitonin Kit

Rapidchip PCT

We created this page for medical staff.

  • *Read carefully the package insert before use.

Reference Data

Within-run Reproducibility

(ng/mL)
  Sample 1 Sample 2
1 0.77 2.2
2 0.62 1.7
3 0.75 1.9
4 0.57 1.8
5 0.66 1.8
6 0.72 2.2
7 0.60 1.9
8 0.67 1.8
9 0.55 2.2
10 0.65 1.8
Mean 0.66 1.9
S.D. 0.07 0.19
C.V.(%) 10.6% 10.0%

Method Comparison

ECLIA vs Rapidchip PCT
CLEIA vs Rapidchip PCT

Correlation(Whole blood vs Plasma)