Digoxin Kit
DRI Digoxin

*Read carefully the package insert before use.

INDEX

Reference Data

Precision (within-run)

(ng/mL)

  N Mean S.D. C.V. (%) Min. Max.
Sample 1 20 0.80 0.02 2.81 0.7 0.8
Sample 2 20 1.99 0.05 2.47 1.9 2.1
Sample 3 20 3.16 0.05 1.62 3.1 3.2

Correlation

  • Measurement range: 0.1-5ng/mL
  • Concentration range for therapy: 0.9-2ng/mL*

Interference

(ng/mL)

Spiked substance Concentration Measurement
Base serum Added serum
Free bilirubin 20mg/dL 2.0 1.9
Conjugated bilirubin 20mg/dL 2.0 2.0
Hemoglobin 500mg/dL 2.0 2.0
Lipemia (Formazin turbidity unit) 3,000 1.9 2.0

Cross reactivity

(ng/mL)

Substance Concentration (ng/mL) Cross reaction rate (%)
Digoxigenin 50 3.6
Digoxigenin bis-digitoxoside 5 108.0
Digoxigenin mono-digitoxoside 5 82.0
Digitoxin 50 4.4
Digitoxigenin 500 0.7
D-(+)-Digitoxose 10,000 0.0
Gitoxin 150 0.5

The following substances didn't show cross-reactivity to the measurement of Digoxin at the following concentrations.

(µg/mL)

Substance Concentration
Acetaminophen 1,000
Cortisone 10
Estriol 10
Phenitoin 100
Prednisone 10
Progesterone 10
Secobarbital 100
Testosterone 5
17-α-OH Progesterone 10
Quinidine 100
Hydrochlorothiazide 100

(µg/mL)

Substance Concentration
Cortisol 10
Dehydroisoandrosterone 10
Lidocaine 100
Prednisolone 10
Propranolol 100
Spironolactone 10
Phenobarbital 100
Procaine amide 100
11-OH-Progesterone 10
Ouabain octahydrate 0.15
Furosemide 50

References