HOME > Description of Business > Diagnostics Business > TDM (drug level in blood) > DRI Digoxin > Reference Data
Digoxin Kit
DRI Digoxin
![]()
- INDEX
-
- Products Top Page
- Reference Data
Reference Data
Precision (within-run)
(ng/mL)
| N | Mean | S.D. | C.V. (%) | Min. | Max. | |
|---|---|---|---|---|---|---|
| Sample 1 | 20 | 0.80 | 0.02 | 2.81 | 0.7 | 0.8 |
| Sample 2 | 20 | 1.99 | 0.05 | 2.47 | 1.9 | 2.1 |
| Sample 3 | 20 | 3.16 | 0.05 | 1.62 | 3.1 | 3.2 |
Correlation

- Measurement range: 0.1-5ng/mL
- Concentration range for therapy: 0.9-2ng/mL*
Interference
(ng/mL)
| Spiked substance | Concentration | Measurement | |
|---|---|---|---|
| Base serum | Added serum | ||
| Free bilirubin | 20mg/dL | 2.0 | 1.9 |
| Conjugated bilirubin | 20mg/dL | 2.0 | 2.0 |
| Hemoglobin | 500mg/dL | 2.0 | 2.0 |
| Lipemia (Formazin turbidity unit) | 3,000 | 1.9 | 2.0 |
Cross reactivity
(ng/mL)
| Substance | Concentration (ng/mL) | Cross reaction rate (%) |
|---|---|---|
| Digoxigenin | 50 | 3.6 |
| Digoxigenin bis-digitoxoside | 5 | 108.0 |
| Digoxigenin mono-digitoxoside | 5 | 82.0 |
| Digitoxin | 50 | 4.4 |
| Digitoxigenin | 500 | 0.7 |
| D-(+)-Digitoxose | 10,000 | 0.0 |
| Gitoxin | 150 | 0.5 |
The following substances didn't show cross-reactivity to the measurement of Digoxin at the following concentrations.
(µg/mL)
| Substance | Concentration |
|---|---|
| Acetaminophen | 1,000 |
| Cortisone | 10 |
| Estriol | 10 |
| Phenitoin | 100 |
| Prednisone | 10 |
| Progesterone | 10 |
| Secobarbital | 100 |
| Testosterone | 5 |
| 17-α-OH Progesterone | 10 |
| Quinidine | 100 |
| Hydrochlorothiazide | 100 |
(µg/mL)
| Substance | Concentration |
|---|---|
| Cortisol | 10 |
| Dehydroisoandrosterone | 10 |
| Lidocaine | 100 |
| Prednisolone | 10 |
| Propranolol | 100 |
| Spironolactone | 10 |
| Phenobarbital | 100 |
| Procaine amide | 100 |
| 11-OH-Progesterone | 10 |
| Ouabain octahydrate | 0.15 |
| Furosemide | 50 |
References
*Jogestrand T,et.al:Am.Heart J. 117, 1076 (1989)
- TDM (drug level in blood) Product List
We prepared this page for medical staff (doctors, dentists, pharmacists, clinical laboratory technicians, nurses, etc.) in Japan to ensure the correct use of our products. You should be aware that the information provided is not intended for overseas medical staff and the general public.