Develops, manufactures, and Distributes various types of in-vitro diagnostics and plastic vacuum blood collection tubes. Diagnostics Business of SEKISUI MEDICAL.

To Previous Page
We created this page for medical staff

Valproic Acid Kit
Nanopia®TDM Valproic Acid

*Read carefully the package insert before use.

INDEX

Reference Data

Precision (within-run)

  Sample 1 Sample 2 Sample 3
No. Sample name BIORAD TDM
Control Level-1
BIORAD TDM
Control Level-2
BIORAD TDM
Control Level-3
Lot No. 57151 57152 57153
1 33.9 66.0 109.3
2 32.2 65.9 109.8
3 34.0 64.9 108.8
4 34.6 65.6 108.8
5 31.4 65.8 108.2
6 31.6 65.9 109.3
7 32.7 65.1 108.4
8 32.7 64.8 108.6
9 33.6 66.1 109.4
10 30.9 65.0 109.7
11 32.2 65.7 107.3
12 29.4 66.6 109.6
13 34.3 69.9 108.8
14 33.8 66.3 109.1
15 34.0 67.5 108.7
16 32.4 67.5 112.5
17 35.2 65.6 108.8
18 34.1 65.4 110.1
19 32.4 67.2 108.9
20 32.6 65.3 110.8
n 20 20 20
Mean 32.90 66.25 109.24
S.D. 1.42 1.19 1.06
C.V. (%) 4.32 1.80 0.97
Max. 35.2 69.9 112.5
Min. 29.4 64.8 107.3
Range 5.8 5.1 5.2

Correlation

Linearity

Interference