Fibrin Degradation Product Kit

Rapidchip D-dimer

We created this page for medical staff.

  • *Read carefully the package insert before use.

Reference Data

Within-run Reproducibility

(µg/mL)
  Sample 1 Sample 2
1 0.55 7.8
2 0.54 6.7
3 0.59 7.4
4 0.52 7.4
5 0.58 7.2
Mean 0.56 7.3
S.D. 0.03 0.4
C.V. 5.2% 5.5%

Method Comparison

LTIA(Plasma) vs Rapidchip D-dimer(Plasma)
LTIA(Whole blood) vs Rapidchip D-dimer(Whole blood)

"Rapidpia" and all related logos are trademarks or registered trademarks of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.
"Rapidchip" is trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.

We prepared this page for medical staff (doctors, dentists, pharmacists, clinical laboratory technicians, nurses, etc.) in Japan to ensure the correct use of our products. You should be aware that the information provided is not intended for overseas medical staff and the general public.