C-Reactive Protein Kit

Nanopia CRP

We created this page for medical staff.

  • *Read carefully the package insert before use.

Reference Data

Diagram

Diagram

Within-run Reproducibility

(mg/dL)
Sample 1 Sample 2 Sample 3
n 20 20 20
Mean 0.1055 0.2114 8.1719
S.D. 0.0019 0.0028 0.0542
C.V.(%) 1.81 1.31 0.66
Range 0.007 0.01 0.23

Linearity

Linearity

Correlation

Correlation

Prozone

Prozone

Interference

Spiked Concentration Measured Value (mg/dL)
Base serum Spiked Serum
Hemoglobin 500mg/dL 0.37 0.36
Unconjugated Bilirubin 50mg/dL 0.36 0.37
Conjugated Bilirubin 50mg/dL 0.37 0.37
Ascorbic Acid 50mg/dL 0.37 0.37
Lipid Emulsion 5% 0.36 0.36

Calibration curve

Calibration curve

Precision

Precision

Detection limit

Detection limit

"Nanopia" is a trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.