Measurement of the serum or plasma soluble interleukin-2 receptor (sIL-2R) concentration
Nanopia IL-2R
We created this page for medical staff.
- *Read carefully the package insert before use.
Reference Data (In-house data of SEKISUI MEDICAL)
On Hitachi 7180
Within-run Reproducibillty
Measurement Value(U/mL) | ||
---|---|---|
Level 1 | Level 2 | |
n | 10 | 10 |
Mean | 531 | 2150 |
S.D. | 8.5 | 13.3 |
C.V.(%) | 1.6 | 0.6 |
Max. | 542 | 2179 |
Min. | 515 | 2136 |
On-board Stability
Detection Limit
Prozone
Interference
Inteference | Spiked Concentration | Measured Value (U/mL) | |
---|---|---|---|
Base sample | Spiked sample | ||
Conjugated Bilirubin | 20mg/dL | 934 | 931 |
Unconjugated Bilirubin | 20mg/dL | 932 | 946 |
Chyle | 3000FTU | 939 | 934 |
Rheumatoid factor | 500IU/mL | 1058 | 1065 |
Hemolytic hemoglobin | 500mg/dL | 935 | 917 |
Correlation
Measurement range (Hitachi 7180)
50-10,000 U/mL
"Nanopia" is a trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.
We prepared this page for medical staff (doctors, dentists, pharmacists, clinical laboratory technicians, nurses, etc.) in Japan to ensure the correct use of our products. You should be aware that the information provided is not intended for overseas medical staff and the general public.