Measurement of the serum or plasma soluble interleukin-2 receptor (sIL-2R) concentration

Nanopia IL-2R

We created this page for medical staff.

  • *Read carefully the package insert before use.

Reference Data (In-house data of SEKISUI MEDICAL)

On Hitachi 7180

Within-run Reproducibillty

Measurement Value(U/mL)
Level 1 Level 2
n 10 10
Mean 531 2150
S.D. 8.5 13.3
C.V.(%) 1.6 0.6
Max. 542 2179
Min. 515 2136

On-board Stability

On-board Stability

Detection Limit

Detection Limit

Prozone

Prozone

Interference

Inteference Spiked Concentration Measured Value (U/mL)
Base sample Spiked sample
Conjugated Bilirubin 20mg/dL 934 931
Unconjugated Bilirubin 20mg/dL 932 946
Chyle 3000FTU 939 934
Rheumatoid factor 500IU/mL 1058 1065
Hemolytic hemoglobin 500mg/dL 935 917

Correlation

Correlation

Measurement range (Hitachi 7180)

50-10,000 U/mL