Digoxin Kit

Nanopia TDM Digoxin

We created this page for medical staff.

  • *Read carefully the package insert before use.

Performance Data

Within-run Reproducibility

  Measured Value
Sample 1 Sample 2 Sample 3
n 10 10 10
Mean 0.72 1.85 3.81
S.D. 0.02 0.02 0.02
C.V.(%) 2.85 0.85 0.46
Max. 0.75 1.87 3.84
Min. 0.68 1.82 3.79

Linearity

Linearity

Correlation

Correlation

Interference

Interfering Substance Spiked Concentration Measured Value (ng/mL)
Base Sample Spiked Sample
Unconjugated Bilirubin 20mg/dL 1.4 1.5
Conjugated Bilirubin 20mg/dL 1.3 1.3
Hemoglobin 500mg/dL 1.4 1.3
Chyle 2500 FTU 1.3 1.9
Ascorbic Acid 50mg/dL 1.3 1.3

Measurement Range(Hitachi 7170S)

0.2-5ng/mL