Digoxin Kit
Nanopia TDM Digoxin
We created this page for medical staff.
- *Read carefully the package insert before use.
Performance Data
Within-run Reproducibility
Measured Value | |||
---|---|---|---|
Sample 1 | Sample 2 | Sample 3 | |
n | 10 | 10 | 10 |
Mean | 0.72 | 1.85 | 3.81 |
S.D. | 0.02 | 0.02 | 0.02 |
C.V.(%) | 2.85 | 0.85 | 0.46 |
Max. | 0.75 | 1.87 | 3.84 |
Min. | 0.68 | 1.82 | 3.79 |
Linearity

Correlation

Interference
Interfering Substance | Spiked Concentration | Measured Value (ng/mL) | |
---|---|---|---|
Base Sample | Spiked Sample | ||
Unconjugated Bilirubin | 20mg/dL | 1.4 | 1.5 |
Conjugated Bilirubin | 20mg/dL | 1.3 | 1.3 |
Hemoglobin | 500mg/dL | 1.4 | 1.3 |
Chyle | 2500 FTU | 1.3 | 1.9 |
Ascorbic Acid | 50mg/dL | 1.3 | 1.3 |
Measurement Range(Hitachi 7170S)
0.2-5ng/mL