Everolimus Kit

Nanopia TDM Everolimus

We created this page for medical staff.

  • *Read carefully the package insert before use.

Performance Data

On Hitachi 7180(except for the correlation)

Within-run Reproducibility

  Measured Value (ng/mL)
Sample 1 Sample 2 Sample 3
n 10 10 10
Mean 4.0 8.0 16.7
SD 0.2 0.29 1.13
CV(%) 5.0 3.6 6.8
Max. 4.2 8.4 18.0
Min. 3.6 7.5 14.5

Detection Limit

Detection Limit

Prozone

Prozone

Correlation

Correlation
(TBA-200FR)
  • *Reference: Soryukai Medical Corporation, Inoue Hospital

Interference

  Spiked Concentration Measured Value (ng/mL)
Base Sample Spiked Sample
Unconjugated Bilirubin 18.7mg/dL 4.0 3.8
Conjugated Bilirubin 19.7mg/dL 4.3 4.3
Rheumatoid factor 360IU/mL 3.9 4.6
Chyle 1440 FTU 4.0 4.1
Intralipos 1.0% 4.3 4.0

Measurement Range

2.0-20.0ng/mL