Performance Data

Within-run Reproducibility

Sample 1 Sample 2 Sample 3
n 20 20 20
Mean 10.19 25.14 51.88
S.D. 0.15 0.51 1.95
C.V.(%) 1.44 2.01 3.76
Max. 10.6 26.3 54.7
Min. 9.9 24.3 48.2
Range 0.7 2.0 6.5

Linearity

Linearity

Correlation

Correlation

Interference

1. Unconjugated bilirubin
2. Conjugated bilirubin
3. Hemoglobin
4. Chyle
5. Ascorbic acid

Measurement Range(Hitachi 7170S)

0.8-80µg/mL

"Nanopia" is a trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.