Vancomycin Kit

Nanopia TDM Vancomycin

We created this page for medical staff.

  • *Read carefully the package insert before use.

Performance Data

On Hitachi 7180(except for the correlation)

Within-run Reproducibility

  Measured Value (µg/mL)
Level 1 Level 2 Level 3
Known Concentration 12.8 38.1 76.9
n 10 10 10
Mean 11.6 34.0 78.3
S.D. 0.15 0.30 5.08
C.V.(%) 1.3 0.9 6.5
Max. 11.5 34.3 89.7
Min. 11.1 33.5 73.0

Detection Limit

Detection Limit

Prozone

Prozone

Correlation

Correlation
(BM6010)

Interference

  Spiked Concentration Measured Value (µg/mL)
Base Sample Spiked Sample
Unconjugated Bilirubin 18.7mg/dL 13.8 13.6
Conjugated Bilirubin 19.7mg/dL 13.7 13.8
Chyle 1440 FTU 13.9 13.7
Hemoglobin 500mg/dL 14.0 14.0
Rheumatoid factor 450IU/mL 14.5 14.6
Intralipos 1.0% 13.6 13.5

Measurement Range

2.5-100µg/mL

"Nanopia" is a trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.