Cholestest N Calibrator

We created this page for medical staff.

Transferability

HDL,LDL T-CHO,TG
  1. Primary Standard (fresh human serum)
    Value assigned by an enzymatic method based on the CDC reference method
  2. Secondary Standard
    (Lyophilized human serum, internal reference material)
    HDL-C
    Value assigned by Cholestest N HDL using the primary standard
    LDL-C
    Value assigned by Cholestest LDL using the primary standard
  1. T-CHO
    Primary Standard (NIST SRM 911)
    TG 
    Primary Standard
    (Value assigned on fresh human serum by JSCC method)
  2. Secondary Standard
    (Lyophilized human serum, internal reference material)
    T-CHO 
    Value assigned by the enzymatic method using the primary standard
    TG
    Value assigned by the enzymatic method using the primary standard
    (free glycerol elimination method)

Cholestest N Calibrator (lyophilized human serum)

Measurement Scheme
HDL-C Value assigned by Cholestest N HDL using the secondary standard
LDL-C Value assigned by Cholestest LDL using the secondary standard
T-CHO Value assigned by the enzymatic method using the secondary standard
TG Value assigned by the enzymatic method using the secondary standard