Cystatin C kit

NORUDIA CYSTATIN C

We created this page for medical staff.

  • *Read carefully the package insert before use.

Reference Data (on Hitachi 7170)

Assay conditions
Sample: 2.4µL
Cystatin C Buffer Solution 1 120µL
Cystatin C Latex Reagent 2 120µL

Within run reproducibility

Measured Value (mg/L)
Low Concentration High Concentration
n 10 10
Mean 0.492 3.941
S.D. 0.004 0.045
CV (%) 0.85 1.15
Max. 0.50 4.04
Min. 0.49 3.89

Detection Limit

Detection Limit

Prozone

Prozone

Interference

Spiked Concentration Measured Value (mg/L)
Base sample Spiked Sample
Unconjugated Bilirubin 20mg/dL 0.68 0.68
Conjugated Bilirubin 20mg/dL 0.68 0.68
Chyle 2000 FTU 0.67 0.69
Hemoglobin 500mg/dL 0.68 0.67
Rheumatoid Factor 500IU/mL 0.75 0.75

On-board Stability

On-board Stability

Correlation

Correlation

Measurement Range:

0.1-10mg/L

Reference Interval:

0.5-1.0mg/L