Albumin kit

NORUDIA U-ALB

We created this page for medical staff.

  • *Read carefully the package insert before use.

Reference Data (on Hitachi 7170)

Within-run Reproducibility

  Measured Value (µg/mL)
Low Concentration Medium Concentration High Concentration
1 11.2 29.4 97.1
2 10.9 29.3 97.9
3 10.5 29.5 102.3
4 11.1 29.2 100.0
5 10.7 29.6 98.4
6 10.9 28.7 99.5
7 10.8 29.5 100.3
8 11.0 29.4 98.6
9 10.9 29.4 98.7
10 11.1 28.5 96.4
Mean 10.91 29.25 98.92
S.D. 0.21 0.36 1.70
CV (%) 1.9 1.2 1.7
Max. 11.2 29.6 102.3
Min. 10.5 28.5 96.4

Linearity

Linearity

Detection Limit

Detection Limit

Prozone

Prozone

Correlation

Correlation

Interference

  Spiked Concentration Measured Value (µg/mL)
Base Sample Spiked Sample
Unconjugated Bilirubin 10mg/dL 38.0 35.6
Conjugated Bilirubin 10mg/dL 38.0 36.9
Chyle 2000 FTU 37.4 38.5
Ascorbic Acid 500mg/dL 82.2 78.5
Hemoglobin 500mg/dL 82.5 85.1
Glucose 4000mg/dL 82.6 82.1
Urea 400mg/dL 80.1 80.8
Chloroform 1.0mL/dL 83.8 83.9
Formalin 1.0mL/dL 83.3 82.7
NaCl 2000mg/dL 43.5 44.4
KCl 1000mg/dL 43.1 43.7

Measurement Range

5-500µg/mL

"NORUDIA" is a trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.