Prothrombin Time Kit
Coagpia PT-Liquid
Intended for medical staff
- *Read package insert carefully before use.
Performance Data (CP3000TM)
Within-run Precision
sec | % | INR | ||
---|---|---|---|---|
Control P-NⅠfor Coagpia | Mean | 12.4 | 101.0 | 1.00 |
S.D. | 0.05 | 0.82 | 0.01 | |
C.V. (%) | 0.39 | 0.81 | 0.48 | |
Max. | 12.4 | 102.2 | 1.00 | |
Min. | 12.3 | 100.5 | 0.99 | |
Range | 0.1 | 1.7 | 0.01 | |
Control P-NⅡfor Coagpia | Mean | 21.8 | 38.9 | 1.79 |
S.D. | 0.40 | 1.01 | 0.03 | |
C.V. (%) | 1.86 | 2.60 | 1.93 | |
Max. | 22.7 | 40.2 | 1.87 | |
Min. | 21.3 | 36.7 | 1.75 | |
Range | 1.4 | 3.5 | 0.12 |
On-Board Stability

Interference of Unfractionated Heparin

Correlation
Correlation between Coagpia PT-N and Coagpia PT-Liquid