Insulin Kit

NORUDIA Insulin

We created this page for medical staff.

  • *Read carefully the package insert before use.

Performance Data

On Hitachi 7180

Within-run Reproducibility

  Measured Value (µU/mL)
Level 1 Level 2 Level 3
n 10 10 10
Mean 6.8 20.4 120.0
S.D. 0.07 0.09 1.42
CV(%) 0.99 0.47 1.18
Max. 6.9 20.5 122.2
Min. 6.7 20.2 118.0

Detection Limit

Detection Limit

Prozone

Prozone

Correlation

Correlation

Interference

  Spiked Concentration Measured Value (µU/mL)
Base Sample Spiked Sample
Conjugated Bilirubin 20mg/dL 12.7 12.7
Unconjugated Bilirubin 20mg/dL 12.7 12.9
Chyle 2000 FTU 12.6 13.3
Intralipos 5% 12.2 12.0
Rheumatoid Factor 500IU/mL 16.8 18.6
Gastrin 10µg/mL 14.2 14.2
Glucagon 10µg/mL 14.2 13.9
Secretin 10µg/mL 14.2 13.4
Proinsulin 12.2ng/mL 0.0 0.0
C peptide 3.9ng/mL 0.0 0.0

On-board Stability

On-board Stability

Measurement Range (Hitachi 7180)

1-150µU/mL

Reference Interval

Fasting insulin level: 5-10 µU/mL

"NORUDIA" is a trademark or registered trademark of SEKISUI MEDICAL CO., LTD. in Japan and/or other countries.