Reference Data

Within-run reproducibility

(µg/mL)
  Sample 1 Sample 2
1 3.5 9.5
2 3.6 9.5
3 3.5 9.4
4 3.4 9.3
5 3.5 9.5
6 3.5 9.3
7 3.5 9.4
8 3.6 9.5
9 3.5 9.5
10 3.5 9.3
Mean 3.5 9.4
S.D. 0.06 0.09
C.V. (%) 1.6 1.0
Max. 3.6 9.5
Min. 3.4 9.3
Range 0.2 0.2

On-Board Stability

[on Coapresta 2000]

On-Board Stability

Interference

  Concentration Measured Value(µg/mL)
Base Plasma Spiked Plasma
Unconjugated Bilirubin 20 mg/dL 2.70 2.80
Conjugated Bilirubin 20 mg/dL 2.80 2.80
Hemoglobin 500 mg/dL 2.12 2.14
Chyle 2500 FTU 2.17 2.04

Linearity

Linearity

Assay Procedure

Assay Procedure