Fibrin Degradation Product Kit
Nanopia D-dimer
We created this page for medical staff.
- *Read carefully the package insert before use.
Reference Data
Within-run reproducibility
Sample 1 | Sample 2 | |
---|---|---|
1 | 3.5 | 9.5 |
2 | 3.6 | 9.5 |
3 | 3.5 | 9.4 |
4 | 3.4 | 9.3 |
5 | 3.5 | 9.5 |
6 | 3.5 | 9.3 |
7 | 3.5 | 9.4 |
8 | 3.6 | 9.5 |
9 | 3.5 | 9.5 |
10 | 3.5 | 9.3 |
Mean | 3.5 | 9.4 |
S.D. | 0.06 | 0.09 |
C.V. (%) | 1.6 | 1.0 |
Max. | 3.6 | 9.5 |
Min. | 3.4 | 9.3 |
Range | 0.2 | 0.2 |
On-Board Stability
[on Coapresta 2000]
Interference
Concentration | Measured Value(µg/mL) | ||
---|---|---|---|
Base Plasma | Spiked Plasma | ||
Unconjugated Bilirubin | 20 mg/dL | 2.70 | 2.80 |
Conjugated Bilirubin | 20 mg/dL | 2.80 | 2.80 |
Hemoglobin | 500 mg/dL | 2.12 | 2.14 |
Chyle | 2500 FTU | 2.17 | 2.04 |
Linearity
Assay Procedure